Line 13: Line 13:
   * Wipe off and decontaminate equipment surfaces and source/destination plates with 70% ethanol before and after use. **Do not use bleach.**   * Wipe off and decontaminate equipment surfaces and source/destination plates with 70% ethanol before and after use. **Do not use bleach.**
   * Ensure source plates are filled to the recommended minimum volume to allow acoustic ejection (typically ≥ 6 µL per well for 384-well LDV plates).   * Ensure source plates are filled to the recommended minimum volume to allow acoustic ejection (typically ≥ 6 µL per well for 384-well LDV plates).
-  * Inspect source plates for bubbles or particulates before loading — these can interfere with droplet ejection.+  * Inspect source plates for bubbles or particulates before loading as these can interfere with droplet ejection.
  
 ---- ----
Line 25: Line 25:
 **User Notes** **User Notes**
  
-  * Always run a **fluid survey** of source plates before performing transfers to verify adequate fluid levels and detect any problematic wells. 
-  * Make sure to **save your transfer protocol** before running as unsaved protocols will be lost if the software closes unexpectedly. 
-  * Review the transfer report generated after each run to confirm all transfers completed successfully. 
  
 ---- ----