====== Beckman Coulter Echo 525 Liquid Handler ====== {{ ::liquid-handler-echo-525-2019-08.webp |}} **About**\\ The Echo 525 Acoustic Liquid Handler is a non-contact, acoustic droplet ejection (ADE) instrument capable of transferring nanoliter to microliter volumes of liquid with high precision and minimal waste. Unlike traditional pipette-based liquid handlers, the Echo 525 uses focused acoustic energy to eject droplets as small as 2.5 nL directly from a source plate to a destination plate without tips. This eliminates cross-contamination, reduces reagent consumption, and enables highly flexible compound and assay dispensing. It is compatible with a wide range of plate formats (384-well, 1536-well) and is controlled by the **Echo Liquid Handler Software**. **User Notes** * Wipe off and decontaminate equipment surfaces and source/destination plates with 70% ethanol before and after use. **Do not use bleach.** * Ensure source plates are filled to the recommended minimum volume to allow acoustic ejection (typically ≥ 6 µL per well for 384-well LDV plates). * Inspect source plates for bubbles or particulates before loading as these can interfere with droplet ejection. ---- ====== Reference Documentation ====== User manual: {{ :echo_525_manual.pdf |}}\\ Consumables: {{ :echo_acoustic_liquid_handler_consumables_brochure.pdf |}}\\ Plate audit applications: {{ :plate_audit.pdf |}}